Name Description Temp Range FI* Config. Mfgr
UpUp Carbon-Ceramic
CCS-ACarbon-Ceramic band A1.5K to 300KUR31K10UATemati
CCS-BCarbon-Ceramic band B2.5K to 300KUR31K10UATemati
CCS-CCarbon-Ceramic band C4.2K to 300KUR31K10UATemati
CCS-DCarbon-Ceramic band D4.2K to 300KUR31K10UATemati
UpUp Cernox™
CX-1010Cernox™0.2 to 325KUACRLakeshore
CX-1030-HTCernox™0.3 to 420KUACRLakeshore
CX-1050-HTCernox™1.4 to 420KUACRLakeshore
CX-1070-HTCernox™4.0 to 420KUACRLakeshore
CX-1080-HTCernox™20 to 420KUACRLakeshore
UpUp Diode
CD-12ASilicon Diode1.4 to 500K4DiodeCryo Industries
CTISilicon Diode10 to 300K8DiodeCTI (Obsolete)
DT-414Silicon Diode1.4 to 500K3DiodeLakeshore
DT-421Silicon Diode1.4 to 500K3DiodeLakeshore
DT-470Silicon Diode1.4 to 500K3DiodeLakeshore
DT-471Silicon Diode1.4 to 500K3DiodeLakeshore
DT-670Silicon Diode1.4 to 500K2DiodeLakeshore
GA-300GaAs Diode25 to 500KUDiodeScientific Instruments
S800Silicon Diode1.4 to 400K6DiodeCryo-con
S900Silicon Diode1.4 to 500K7DiodeCryo-con
SI410Silicon Diode1.4 to 500K5DiodeScientific Instruments
SI430Silicon Diode1.4 to 500K7DiodeScientific Instruments
SI440Silicon Diode1.4 to 500K7DiodeScientific Instruments
TG-120GaAs Diode25 to 500KUDiodeLakeshore
UpUp Platinum RTD
CP-100100 Ohm Platinum RTD14 to 873K20PTC100Cryo-con
GP-100100 Ohm Platinum RTD14 to 673K20PTC100Cryo-con
PT100 RTDGeneric 100 Ohm Platinum RTD20 to 800K20PTC100generic
PT-102100 Ohm Platinum RTD14 to 873K20PTC100Lakeshore
PT10K RTDGeneric 10K Ohm Platinum RTD20 to 800K22PTC10Kgeneric
PT-111100 Ohm Platinum RTD14 to 673K20PTC100Lakeshore
PT1K RTDGeneric 1,000 Ohm Platinum RTD20 to 800K21PTC1Kgeneric
XP-100100 Ohm Platinum RTD14 to 425K20PTC100Cryo-con
XP-1K1K Ohm Platinum RTD14 to 425K21PTC1KCryo-con
UpUp Rhodium-Iron RTD
PHZ 0002Rhodium-Iron RTD1.4 to 500K23PTC100Oxford
RF-800Rhodium-Iron RTD1.4 to 500K23PTC100Lakeshore
UpUp Ruthenium-Oxide
R400Ruthenium-Oxide1.4 to 40K34ACRCryo-con
RO-105Ruthenium-Oxide1.4 to 40K32ACRScientific Instruments
RX-103ARuthenium-Oxide1.4 to 40KUACRLakeshore
*FI = Factory Installed. 0 to 60 indicates that the sensor is supported by a factory installed sensor calibration curve.      U indicates that support in temperature units requires a user supplied curve.
Input configurations are:
 R250K10UA - DC resistance. Constant-current excitation is 10uA. Full-scale resistance is 250KΩ.
 R31K10UA - DC resistance. Constant-current excitation is 10uA. Full-scale resistance is 31KΩ.
 Diode - DC Diode. Excitation 10uA DC. Input voltage range: 2.25V
 PTC100 - DC resistance. Excitation 1.0mA DC. Full scale resistance is 625Ω.
 PTC1K - DC resistance. Excitation 1.0mA DC. Full scale resistance is 625Ω.
 PTC10K - DC resistance. Excitation 1.0mA DC. Full scale resistance is 625Ω.