Cryocon Navigation Bar
Products Support Company Home Distributors Home

 Home > Temperature Monitors > Model 18 > Specifications

Model 18 Cryogenic Temperature Controllers

Specifications

 

UpUp Model 18 General Specifications

User Interface

Display Type: Graphics VFD, 4mm character height.

Number of Inputs Displayed: Eight.

Keypad: Sealed Silicon Rubber.

Temperature Display: Six significant digits, autoranged.

Display Update Rate: 0.5 Seconds.

Display Units: K, C, F or native sensor units.

Display Resolution: Up to seven significant digits.

Input Channels

There are eight identical input channels, each may be independently configured for any of the supported sensor types.

Sensor Connection: 4-wire differential. DB-9 receptacle.

Sensor Types: See supported sensors table.

Sensor Selection: Front Panel or remote interface.

Sensor Excitation: DC constant-current: 1mA, 100µA or 10µA.

Resistance Measurement type: Ratiometric bridge.

Resistance Range: 1Ω to 250KΩ.

Sample Rate: 0.8Hz per channel.

Digital Resolution: 24 bits.

Digital Accuracy: 0.0015% of full scale.

Measurement Drift: <25ppm/oC.

Measurement Filter: 0.5, 1, 2, 4, 8, 16, 32 and 64 Seconds.

Calibration Curves: Built-in curves for industry standard sensors plus four user curves with up to 200 entries each. Interpolation is performed using a Cubic Spline.

Data Logging

Data logging is performed to an internal, 40K byte circular buffer and is time-stamped with a real-time clock. Buffer memory is non-volatile and will retain valid data without AC power. The Model 18 will log a maximum of 1,000 entries where each entry contains eight temperature readings.

Status Indicators

Alarm LED on the front panel indicates a programmable alarm condition based on input temperature.

Remote Interfaces

Ethernet: Connects to any 10/100 Base-T Ethernet type Local Area Network. Electrically isolated. Connection and activity indicators on rear panel. IP address and sub-net mask are user configurable. TCP/IP user data socket provides remote control and interface to common data acquisition software by using an ASCII command language. HTTP provides built-in web server. SMTP sends e-mail based on user selected alarm conditions.

Language: Remote interface language is IEEE SCPI compliant.

LabView™ Drivers are available.

General

Ambient Temperature: 25°C ± 5 oC for specified accuracy.

Mechanical: 5.6”W x 2.9”H x 8.8”D.

Weight: 3.5 Lbs.

AC Power Requirement: 110 or 220VAC, +5% to -10%, 50 to 60Hz, 10W.

AC Power Switch: Rear panel.

Conformity: European CE certified.

Calibration: NIST traceable.

UpUp Model 18Supported Sensors

Silicon Diode sensors use a fixed excitation current of 10µA and an input voltage range of 0 to 2.25V.

Negative and Positive Temperature Coefficient (NTC) / (PTC) resistor sensors are supported using a ratiometric bridge technique to cancel low frequency noise.

Platinum RTD sensors can use built-in DIN 43760 (IEC 750) standard setups for 100Ω or 1,000Ω devices. The Model 18 uses the DIN standard for temperatures from 70K to 1020K and extends it for cryogenic use. Operation down to 14K is possible with user supplied curves. A unique feature is the use of a ratiometric resistance bridge technique to measure Platinum RTD sensors. This significantly reduces low frequency noise and drift to provide rock-solid measurements.

Negative Temperature Coefficient (NTC) sensors commonly used in cryogenic applications include Ruthenium-Oxide, Cernox™, Carbon Ceramic and several others. These sensors are supported by using a 10μA constant-current DC excitation with a 2.50V or 0.31V input range.

 

 

page Listing of Supported Sensors

Model 18 Supported Sensors


Temperature Range

Example Sensors

Diode

1.4 - 500K

Cryo-con S900, S800
SI-440, 430, 410
Lakeshore DT-670, 470

Platinum RTD

14 - 1200K

Cryo-con CP-100
Cryo-con GP-100
Cryo-con XP-100
Cryo-con XP-1K

Rhodium-Iron

1.4 - 800K

Oxford PHZ 0002

Cernox™

2.0K - 325K

Lakeshore, all types

Carbon-Ceramic

4.2K - 300K

TMi-A1

Ruthenium-Oxide

2.0K - 200K

Cryo-con R400
Cryo-con R500

UpUp Model 18 Input Specifications

Input Specifications


Diode sensors

PTC resistor sensors

NTC resistor sensors

Excitation Mode

DC, Constant-Current

DC, Constant-Current

DC, Constant-Current

Ranges

2.5V

2.5KΩ, 25KΩ, 250KΩ

250KΩ

Accuracy:
 % of Rdg + % of range

0.005% + 0.001%

0.021% + 0.001%

0.021% + 0.001%

Resolution:
  % of Range

0.00004%

0.0003%

0.0003%

Excitation Current

10µA DC

1.0mA, 100mA, 10mA

10µA

CalGen™ and Cryo-Con™ are registered trademarks of Cryogenic Control Systems, Inc. All other product and company names are trademarks
or trade names of their respective companies.
Tel: (858) 756 3900   Fax: (858) 759 3515   
Specifications subject to change without notice. ©Cryogenic Control Systems, Inc. 1998 - 2008.