Silicon diodes are easily the most common temperature sensors used in cryogenic applications. They offer good sensitivity and a wide range of operation compared to other devices. Further, they have excellent interchangeability and stability.
The S900 is offered in three packages to fit a wide range of applications:
For applications above 50K, please consider Cryo-con's Platinum RTDs

Temperature Range: 1.5 to 400K.
Marking: Individual serial number on the top of the package.
Mass: 1.1g excluding leads.
Construction: Bobbin Material is gold plated oxygen free hard copper. Sensor bonding is Stycast® epoxy.
Leads: 36”, 36AWG Phosphor-Bronze. Four-lead color coded cryogenic ribbon cable. Insulation is heavy Formvar® . Strain relief is Teflon®.
Mounting: 4-40 brass machine screw. A thin layer of Apiezon® N grease is recommended.


Temperature range: 1.5 to 400K.
Mass: 37mg
Thermal response time: <10mS at 4.2K, 100mS at 77K and 200mS at 305K.
Construction: Package material is gold plated OHFC copper on an Alumina substrate.
Leads: 3”, 37 AWG copper with Polyimide insulation. Positive connection is red and negative is black. Lead attach is 60/40 solder.
Mounting: Varnish, epoxy or clamp to a clean flat surface. Thermally anchor copper leads.


Temperature Range: 1.4 to 500K
Mass: 32mg
Thermal response time: <10mS at 4.2K, 100mS at 77K and 200mS at 305K.
Construction: Package material is Alumina with gold plated OHFC copper end caps. Leads are 3”, 37 AWG copper with Polyimide insulation.
Mounting: Sensor body must be electrically insulated from mounting surface. Attach using varnish or epoxy. Thermally anchor the copper leads.
Cryocoolers and cryogen free systems.
Temperature Range: 1.5K to 400K for the BB and SM packages. 1.5K to 500K for the CP package.
Standard Curve: Cryo-con S900.
Excitation Current: 10µA ±0.1%
Repeatability:10mK @4.2K, 16mK @77K, 75mK @273K.
Magnetic Field Use: Not recommended for fields above 0.1 Tesla or temperature below 40K.
Use in Radiation: Not recommended for high
radiation environments.
Maximum Temperature: Do not store above 500K.
Maximum Reverse Voltage: 60V.
Maximum Excitation Current: 2mA.
Power Dissipation: 16µW at 4.2K.
1.5K - 30K |
30K - 100K |
100K - 400K |
|
|---|---|---|---|
| Band A | ±0.1K | ±0.2K | ±1.0K |
| Uncalibrated | ±0.2K | ±0.4K | ±1.0K |



T(K) |
V |
S(mV/K) |
|---|---|---|
| 1.5 | 1.63 | -31.5 |
| 4.2 | 1.55 | -38.9 |
| 77.35 | 1.03 | -1.7 |
| 273 | 0.62 | -2.29 |
| 500 | 0.1 | -2.1 |
| CalGen™ and Cryo-Con™ are registered
trademarks of Cryogenic Control Systems, Inc. All other product and company names are trademarks or trade names of their respective companies. |
|
Tel: (858) 756 3900 Fax: (858) 759 3515
Specifications subject to change without notice. ©Cryogenic Control Systems, Inc. 1998 - 2011. |