Cryocon Navigation Bar
Products Support Company Home Distributors Home

 Home > Monitor Selection Guide

Temperature Monitor

Selection Guide


  Model 12
Model 12
Model 14
Model 14
Model 18C
Model 18C
  Two input, Operation to 200mK Four input, Operation to 200mK Eight input, Operation to 500mK
Sensor Inputs
2
4
8
Temperature Range 200mK to 1020K 500mK to 1020K 1.4K to 1020K
Excitation Current 1mA to 32nA AC/DC 1mA to 32nA AC/DC 1mA, 10uA
Constant-Voltage Excitation Levels 1.0V, 1mV, 10mV 1.0V, 1mV, 10mV 100mV / 10mV
Supported Sensors

Silicon Diode

1.4K to 475K 1.4K to 475K 1.4K to 475K

Ruthenium Oxide

200mK to 273K 200mK to 273K 200mK to 273K

Platinum RTD

14K to 1200K 14 to 1200K 14K to 1200K

Cernox™

200mK to 420K 200mK to 420K 200mK to 420K

Rhodium-Iron

1.4K to 800K 1.4K to 800K 1.4K to 800K

Germanium

300mK to 100K 300mK to 100K 300mK to 100K

Carbon Glass™

1.0K to 325K 1.0K to 325K 1.0K to 325K

Thermocouple

N / A N / A Optional. Types E, K, T and AuFe. 1.5K to 1500K
Calibration Curves
Built-in curves for most industry standard sensors plus additional user supplied curves.
   Interpolation is performed using a Cubic Spline algorithm for highest possible accuracy.
User Calibration Curves 4 x 200 entries 4 x 200 entries 8 x 200 entries
 
Computer Interfaces
Ethernet Yes Yes Yes
 
Ethernet interface includes: Http: Embedded Web server for temperature monitoring and
   instrument configuration using any web browser, SMTP: E-mail and TCP/IP: Client for
   ASCII command language remote control.

LabView™ drivers available for all interfaces including Ethernet TCP/IP.

Cryo-con ezNet software drivers provided for quick TCP/IP communication to any
   Windows™ based PC.
IEEE-488.2 No No Option
RS-232 Yes Yes Yes
USB Optional Optional Optional
Alarms 1 - LED 1 - LED Audio / LED
Analog Outputs Zero to 4.096 Volts Zero to 4.096 Volts N/A
Relay Outputs 2 Dry contact, 30W 2 Dry contact, 30W 2 - 10A
 
User Interface
Front Panel Display
VFD. 2 x 10 and 4 x 20 font.
TFT LCD, 8 x 21 or 4 x 10 font.
Number of channels
    displayed
2 4 8
  Keypad
5 key, Sealed Silicon Rubber.
4 key
 
Internal Data Logging
Type Internal, non-volatile Internal, non-volatile Internal, non-volatile
Samples 1,000 500 1,000
 
General
Power Input

110/240 VAC, 50-60V

by internal switch setting

110/240 VAC, 50-60V

by internal switch setting

7.5 - 48V AC or DC or
IEEE-802.3at Power-Over-Ethernet.

External transformer supplied.

Mechanical

1/3 DIN
5.6"W x 2.9"H x8.8"D

1/3 DIN
5.6"W x 2.9"H x8.8"D

1/3 DIN
5.6"W x 2.9"H x8.8"D

CalGen™ and Cryo-Con™ are registered trademarks of Cryogenic Control Systems, Inc.
All other product and company names are trademarks or trade names of their respective companies.
Tel: (858) 756 3900   Fax: (858) 759 3515   
Specifications subject to change without notice. ©Cryogenic Control Systems, Inc. 1998 - 2011.